1
|
- Dónal J Murphy PhD
- Analytical Development
- AstraZeneca UK
|
|
2
|
|
|
3
|
- Method Transfer Considerations
- Tablet drop
- Temperature differential
- Filters
- Automated media preparation
- Equivalence with manual method
- Examples of system application
|
|
4
|
- Tablet drop
- Successful automatic tablet drop can be influenced by tablet shape and
dimensions
- Correct choice of tablet magazine will ensure successful drop
- Tablet magazines of varying hole dimensions are available
|
|
5
|
|
|
6
|
- Temperature Differential
- Standards and samples are not at the same temperature (using standard
equipment)
- Potential for spectral effects
- Check for differences in lmax and Abs
- Apply thermal control if necessary
|
|
7
|
- Cleaning Validation
- Need to ensure cleaning is adequate to prevent carryover
- Perform run using max dose followed by blank run
- Cleaning uses water only
- Possible issues for water insoluble drugs/excipients
- Media containing surfactants – also need to ensure adequate removal of
surfactant
|
|
8
|
- Filters
- Standard configuration works best with 25mm Acrodisc syringe filters
(Pall Gelman)
- Choice of filter limited by back pressure
- Glass fibre and GHP work well
- Maximum number of unattended runs limited by number of filters required
per run
- Could one filter be used for all timepoints?
|
|
9
|
|
|
10
|
- Automated Media Preparation
- Optional extra
- Prepares by dilution from concentrate (x10)
- pH accuracy of ± 0.05 pH units required
- Validate and verify
- Consider pH monitoring option
- Consider temperature effect
|
|
11
|
- Equivalence with manual method
- How many batches to test?
- How many tablets to test?
- Are results comparable?
- Accuracy
- Precision
- What statistical tool(s) to use?
- Evidence of bias?
|
|
12
|
|
|
13
|
|
|
14
|
- Analysis of difference
- Does equivalence hide potential problems?
- Is hollow shaft sampling an issue?
- Do different filters have an effect?
- Sampling & filtration effects?
- Traditional sampling position – do resident cannulae have an effect?
|
|
15
|
|
|
16
|
|
|
17
|
|
|
18
|
|
|
19
|
|
|
20
|
|
|
21
|
|
|
22
|
- What do we use the system for?
- Clearances
- Stability samples
- Development runs (method and formulation)
- Where long runs times are required (e.g. data for IVIVC)
- Investigations where greater precision is required
|
|
23
|
|
|
24
|
|
|
25
|
|